RKI Instruments EAGLE 2 Manual User Manual

Page 115

Advertising
background image

EAGLE 2 Operator’s Manual

Using Setup Mode • 105

2. Press and release POWER ENTER RESET. The Set Date Format screen

appears with the current setting flashing.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Setting the Battery Type

This menu item allows you to select between alkaline and Ni-MH batteries.
Since the discharge characteristics of alkaline and Ni-MH batteries are
different, the EAGLE 2 uses this setting to ensure that the low battery
warning is in effect long enough before a dead battery alarm to allow the user
to change the batteries without a dead battery alarm occurring. This setting
has no effect on battery charging.

1. From the main menu, place the cursor next to SET BATTERY TYPE.

2. Press and release POWER ENTER RESET. The Battery Type screen

appears with the current setting flashing.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Configuring the Channels

This menu item allows you to set the channel type for each of the six
channels or to turn one or more channels off. Although the standard EAGLE
2 is factory configured for four channels, combustible gas (catalytic sensor),
oxygen, H

2

S, and CO, with channels 5 and 6 turned off, the EAGLE 2 can be

factory and field configured for a variety of active channels and detector
types. It is not normally necessary to change the factory channel

SET DATE FORMAT

UP/DOWN THEN ENTER

MM/DD/YYYY

BATTERY TYPE

UP/DOWN THEN ENTER

ALKALINE

Advertising