120 • Using Setup Mode

EAGLE 2 Operator’s Manual

3. Use AIR ▲ YES and RANGE ▼ SHIFT to adjust the time to the desired

setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the Auto Fresh Air Setting

This setting allows you to configure the EAGLE 2 so that a fresh air
adjustment takes place automatically as part of the instrument startup
sequence. If AUTO FRESH AIR ADJ is set to ON, the EAGLE 2 performs
a fresh air adjustment at the end of the startup sequence before entering
Normal Operation. The factory setting is OFF.

WARNING: If the automatic fresh air feature is turned on, you must

startup the EAGLE 2 in a known fresh air environment, an
environment free of toxic or combustible gases and of normal
oxygen content (20.9%). If this feature is on and the EAGLE
2 is started up in the presence of a target gas, the readings
and alarms will not be accurate or reliable.

1. From the main menu, place the cursor next to AUTO FRESH AIR ADJ.

2. Press and release POWER ENTER RESET. The Fresh Air Adjust Screen

appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the Data Log Interval Setting

This setting indicates how often the EAGLE 2 saves readings to the data
logger. The following interval times can be selected: 10 minutes, 5 minutes, 3
minutes, 1 minute, 30 seconds, 20 seconds, 10 seconds, or 5 seconds. The
factory setting is 30 seconds.

1. From the main menu, place the cursor next to DATA LOG INTERVAL.

FRESH AIR ADJUST

ON POWERUP

UP/DOWN THEN ENTER

OFF