Maintenance, Troubleshooting, Cleaning the pid sensor’s lamp – RKI Instruments EAGLE 2 Manual User Manual

Page 170

Advertising
background image

160 • Maintenance

EAGLE 2 Operator’s Manual

Maintenance

The PID sensor includes user serviceable parts. They are the lamp and the
electrode stack. The following sections include troubleshooting instructions
as well as procedures for cleaning the lamp, replacing the lamp, replacing the
electrode stack, and replacing the PID sensor.

Troubleshooting

The troubleshooting table describes symptoms, probable causes, and
recommended action for problems you may encounter with the PID sensor.

Cleaning the PID Sensor’s Lamp

Clean the lamp if you notice a significant drop in sensitivity from one
scheduled calibration to another or if you are not able to calibrate the PID
channel. See the “PID EAGLE 2 Spare Parts” on page 178 for lamp cleaning
kit ordering information. The lamp cleaning kit includes the following items:

an electrode stack removal tool

Symptoms

Probable Causes

Recommended Action

• PID channel

doesn’t exit its
warm up state and
the reading
continues to scroll
through -200, -100,
and 0 for a high
range sensor and
-2.00, -1.00, and
0.00 for a low
range sensor.

• The PID sensor has not

come out of its warm up
state.

1. Clean the lamp.
2. Replace the lamp.
3. Replace the electrode stack.
4. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• The PID sensor

fails to calibrate.

• The auto calibration

values may not match the
cylinder gas concentration
(auto calibration only).

• The sample gas is not

reaching the sensor
because of a bad
connection.

• The calibration cylinder

may be out of gas or is
outdated.

• The lamp may need to be

cleaned.

• The electrode stack may

need to be replaced.

1. Make sure the EAGLE 2 has

been properly set up for
calibration.

2. Check all calibration tubing for

leaks or for any bad connections.

3. Verify that the calibration cylinder

contains an adequate supply of
sample.

4. Clean the lamp.
5. Replace the electrode stack.
6. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

Advertising