EAGLE 2 Operator’s Manual

Display Mode • 41

NO button.

To enter an item when a question is asked, press and release the AIR ▲
YES button.

To change a flashing parameter, use either the AIR ▲ YES button or
RANGE ▼ SHIFT button.

NOTE: Each screen displays for 20 seconds. If you do not press a button

within 20 seconds, the EAGLE 2 automatically returns to Measuring
Mode.

Peak Screen

The peak screen displays the highest (lowest for oxygen) concentrations
detected since the EAGLE 2 was turned on. Peak readings are stored in the
EAGLE 2’s memory until a higher level is detected (lower for oxygen), the
peak reading is cleared, or the EAGLE 2 is turned off.

The lunch break feature enables the EAGLE 2 to save peak readings when it
is turned off so it can continue them when it is turned on again. See “Turning
On the EAGLE 2” on page 22

To clear the peak readings, do the following:

1. With the Peak Screen displayed, press and release the POWER ENTER

RESET button. The following screen will appear.

2. Press and release the AIR ▲ YES button. The peak readings will be reset

and the unit will return to the Peak Screen.

If you do not want to clear the peak readings, press and release the
DISPLAY ADJUST NO button and the unit will return to the Peak
Screen without clearing the peak readings.

CH4 0%LEL
P OXY 20.9vol%
E H2S 0.0ppm
A CO 0ppm
K

CLEAR

PEAK READINGS?