EAGLE 2 Operator’s Manual

PID Relative Response Feature • 141

2. With the PID Sensor Relative Response Screen displayed, press and

release AIR ▲ YES. A list of gases will appear on the screen with EXIT
at the top of the list.

There are multiple screens of gases. The following is the complete list of
factory defined gases along with their detection ranges, low alarm, high
alarm, STEL, and TWA settings. Table 14 is a list of the low range values
and Table 15 is a list of the high range values.

Table 14: Low Range Relative Response Gas List

Target Gas

Detection

Range (ppm)

Alarm 1 Factory

Setting (ppm)

Alarm 2 Factory

Setting (ppm)

STEL

(ppm)

TWA

(ppm)

Acetone

0-30.00

5.00

7.50

OFF

OFF

Benzene

0-25.00

0.50

2.50

2.50

0.50

Diesel Fuel

NO 1

0-40.00

2.00

3.00

OFF

OFF

Ethanol

0-400.00

10.0

15.0

OFF

OFF

Gasoline

0-50.00

3.00

5.00

OFF

OFF

Isobutylene

0-50.00

4.30

6.00

OFF

OFF

Isopropanol

0-200.00

2.0

4.0

OFF

OFF

JP-5 Fuel

0-30.00

1.40

2.10

OFF

OFF

Methyl Ethyl

Ketone

0-40.00

2.00

3.00

OFF

OFF

Toluene

0-25.00

0.50

1.50

OFF

OFF

N-Hexane

0-200.00

5.0

10.0

OFF

OFF

Propylene

0-50.00

5.00

7.50

OFF

OFF

Styrene

0-20.00

0.20

0.40

OFF

OFF

Tetrachloro-

ethylene

0-30.00

0.24

1.00

OFF

OFF

Trichloro-

ethylene

0-30.00

0.50

1.00

OFF

OFF

Vinyl Chloride

0-100.00

1.0

5.0

5.0

1.0