EAGLE 2 Operator’s Manual

Configuring the TC Gas in Setup Mode • 219

8. Use RANGE ▼ SHIFT to move the cursor down the menu to the TC

sensor.

9. To change the TC sensor gas configuration, press and release POWER

ENTER RESET.

10. A screen appears that indicates the detection range for the currently

configured gas. In the example below, the TC sensor is currently
configured for 0 - 100 %volume CH

4

. The screen also shows the gas

configuration choices for the TC sensor. There are 3 pre-defined options
and 1 user-defined option. The user-defined option will always have an
asterisk next to it.

11. Use AIR ▲ YES and RANGE ▼ SHIFT to move the cursor next to the

desired gas.

12. If you placed the cursor next to one of the pre-defined gases, press and

release POWER ENTER RESET to select the gas and proceed to Step 17.

If you placed the cursor next to the user defined gas with the asterisk (*),
press POWER ENTER RESET and proceed with Step 13.

13. The user defined gas setup screen appears with the first character of the

gas name flashing. The current gas name and range are shown on the top
line of the screen.

CONFIGURE GASES

CAT : CH4 (CAT)

> OP1 : CH4 (TC)

OP2 : --- (---)

OP3 : --- (---)

TC (0-100.0vol%CH4)

> 100.0 vol% CH4

100.0 vol% H2

10.0 vol% H2

10.0 vol% TC *

TC (0-100.0vol%CH4)

CHANGE TO NAME

TC

FULL_SCALE 10.0