128 • Using Setup Mode

EAGLE 2 Operator’s Manual

password is saved and you return to the main menu.

Restoring the Default Settings

Each of the EAGLE 2 setup parameters, such as the auto calibration values,
zero and span settings, or parameters in Setup Mode, has a default setting.
For the items in Setup Mode, the default settings are the same as the standard
factory settings. If you want to return the EAGLE 2 to its default
configuration, it is possible to do so by using the Default Settings menu item
in Setup Mode. Returning the EAGLE 2 to its default configuration can be
useful if various setup parameters have been changed in the field and you
want to return the EAGLE 2 to its original configuration as shipped from the
factory.

The standard default gas configuration is LEL/oxygen/H

2

S/CO. If you have

turned any channels off or have added channels to your EAGLE 2, you will
have to re-setup your EAGLE 2 to the desired gas combination if you restore
the EAGLE 2 to its default configuration.

There are some special EAGLE 2 configurations that may have a different
default configuration than the standard. Consult RKI Instruments, Inc. for
information regarding non-standard default configurations.

WARNING: When the EAGLE 2 is restored to its default configuration,

the zero and span values for each channel are reset. You must
recalibrate all active channels if you restore the EAGLE 2 to
its default configuration.

1. From the main menu, place the cursor in front of DEFAULT

SETTINGS.

2. Press and release POWER ENTER RESET. The Restore Default

Configuration? Screen appears asking if you want to restore the default
configuration.

3. If you do not want to restore the default configuration, press and release

DISPLAY ADJUST NO to return to the main menu.

If you do want to restore the default configuration, continue with Step 4.

RESTORE DEFAULT

CONFIGURATION?