EAGLE 2 Operator’s Manual

Measuring Mode, Alarms • 37

Resetting and Silencing Alarms

You can set the EAGLE 2’s gas alarms as latching or self-resetting alarms
(see “Updating the Alarm Latching Setting” on page 116).

Self-resetting alarms (ALARM LATCHING set to SELF RESET)

Self-resetting alarms automatically shut off and reset when the gas reading
falls below (or rises above for an oxygen low alarm) the alarm setting. You
cannot reset self-resetting alarms with the POWER ENTER RESET
button. You can set self-resetting alarms with or without the alarm silence
feature (see “Updating the Alarm Silence Setting” on page 117).

Latching alarms (ALARM LATCHING set to LATCHING)

Latching alarms will remain in effect until the gas reading falls below (or
rises above for an oxygen low alarm) the alarm setting and they are reset
with the POWER ENTER RESET button. You can set latching alarms
with or without the alarm silence feature (see “Updating the Alarm
Silence Setting” on page 117).

ALARM SILENCE On and Alarms Set to LATCHING:

ALARM SILENCE set to ON and ALARM LATCHING set to LATCHING

Low Flow

• The display indicates FAIL

LOW FLOW LEVEL

• Alarm LED arrays flash in a

double pulsing pattern once
per second

• Backlight turns on

Double pulsing tone
once per second

Low Battery Warning

BATT appears vertically along

the left side of LCD

None

Dead Battery Alarm

• Gas readings replaced by

RECHARGE INSTRUMENT
BATTERIES

• Alarm LED arrays flash in a

double pulsing pattern once
per second

Double pulsing tone
once per second

Sensor Failure

FAILED SENSOR(S) appears

at the top of the display and
the failed sensor(s) are
indicated

• Alarm LED arrays flash in a

double pulsing pattern once
per second

Double pulsing tone
once per second

Table 6: Alarm Types and Indications

Alarm Type

Visual Indications

Audible Indication