182 • ESM-01 Calibration

EAGLE 2 Operator’s Manual

described below in “Calibrating with a 5-Gas Cylinder” on page 191, uses a
multi-gas cylinder that contains the target gas for all installed sensors. Follow
the set of instructions appropriate for your cylinder configuration.

NOTE: If your instrument has more than one ESM-01 sensor, you will need

a calibration cylinder for each sensor and follow the instructions in
“Calibrating with a 4-Gas Cylinder and an ESM-01 Cylinder” on
page 182.

Calibrating with a 4-Gas Cylinder and an
ESM-01 Cylinder

The calibration instructions below show a 5 channel instrument which has the
four standard channels, LEL/oxygen/H

2

S/CO, and an ESM-01 channel for

ammonia (NH

3

).

1. See “Calibration Supplies and Equipment” on page 56 for a description

of the necessary calibration supplies. In addition to an appropriate multi-
gas cylinder that is used to calibrate any active standard channels, you
will also need a cylinder to calibrate the ESM-01 channel. See Table 19
on page 202 for available cylinders. Make sure your calibration cylinder
is appropriate for the ESM-01 detection range.

2. Find a fresh-air environment. This is an environment free of toxic or

combustible gases and of normal oxygen content (20.9%).

3. While in Measuring Mode, press and hold the RANGE ▼ SHIFT button,

then press the DISPLAY ADJUST NO button and release both buttons.

4. If the unit prompts you for the password, enter it by using the AIR ▲

YES and RANGE ▼ SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.

5. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION