96 • Calibrating Using the Single Calibration Method

EAGLE 2 Operator’s Manual

Performing a Span Adjustment in Single Calibration

1. Slide the tubing clamp onto the tubing and connect the tubing to the

sample bag’s inlet. Leave the clamp unclamped for now.

2. Connect the other end of the tubing to the regulator’s or dispensing

valve’s hose barb fitting.

3. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is

complete with internal O-ring and membrane and that the two halves of
the probe are tightened firmly together to avoid leaks that can affect the
calibration. See Figure 19, “Replacing the Particle Filter and
Hydrophobic Filter Disk” on page 76 for an illustration of the internal
parts of the probe.

4. Move the cursor next to the SINGLE CALIBRATION menu item by

using the AIR ▲ YES button.

5. Press and release the POWER ENTER RESET button. The Select Sensor

Screen appears with the cursor flashing.

6. Move the cursor next to the sensor you want to calibrate with the AIR ▲

YES and RANGE ▼ SHIFT buttons. In the example below, the CH

4

sensor is selected for span adjustment.

If you do not want to proceed with the span adjustment, press and release
the DISPLAY ADJUST NO button or place the cursor next to ESCAPE

CALIBRATION MODE

AUTO CALIBRATION

> SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

SELECT SENSOR
TO CALIBRATE

>ESCAPE

CH4 OXY
H2S CO

SELECT SENSOR
TO CALIBRATE

ESCAPE

>CH4 OXY

H2S CO