EAGLE 2 Operator’s Manual

Infrared CO

2

Calibration • 229

11. The EAGLE 2 will indicate that it is adjusting the zero reading for a few

seconds, then indicate that the operation is complete before returning to
the Calibration Mode Screen.

12. Remove the CO

2

scrubber from the EAGLE 2 inlet fitting.

13. Put the caps back on the scrubber.

14. Install the demand flow regulator onto the multi-gas calibration cylinder.

15. Connect the sample tubing to the demand flow regulator.

16. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is

complete with internal O-ring and membrane and that the two halves of
the probe are tightened firmly together to avoid leaks that can affect the
calibration. See Figure 19, “Replacing the Particle Filter and
Hydrophobic Filter Disk” on page 76 for an illustration of the internal
parts of the probe.

ADJUSTING ZERO

RELEASE AIR BUTTON

ADJUSTING ZERO

COMPLETE

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION