RKI Instruments EAGLE 2 Manual User Manual

Page 123

Advertising
background image

EAGLE 2 Operator’s Manual

Using Setup Mode • 113

Mode.

13. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired catalytic

sensor voltage, 1.30 volts or 2.40 volts.

14. Press and release POWER ENTER RESET. The confirmation screen

appears. In the example below, the user defined gas has been selected and
defined as propane with the gas name set to PRO.

15. If you want to accept the gas configuration change, press and release AIR

▲ YES. The unit will return to the Configure Gases screen.

If you do not want to accept the gas configuration change, press and
release DISPLAY ADJUST NO to return to the screen with the gas
choices shown in Step 4 on page 108. You can either scroll down to END
and press POWER ENTER RESET to return to the Configure Gases
screen or continue from Step 4 on page 108 to select a new gas.

16. Use RANGE ▼ SHIFT to place the cursor next to END and press

POWER ENTER RESET to return to the main menu.

17. Be sure to perform a calibration on the catalytic combustible channel

using an appropriate calibration cylinder for the target gas.

Setting the Catalytic Detection Units

This menu item allows you to display the combustible gas units on the
catalytic sensor channel as ppm, %LEL, %vol, or selectable between the
three units (the CHANGE OK option). The factory setting is CHANGE
OK
.

1. From the main menu, place the cursor next to CATALYTIC UNITS.

CHANGE TO PRO ?

PRESS YES OR NO

CONFIGURE GASES

> CAT : PRO (CAT)

OP1 : --- (---)
OP2 : --- (---)

OP3 : --- (---)

Advertising