EAGLE 2 Operator’s Manual

Calibrating Using the Auto Calibration Method • 89

2. Press and release the POWER ENTER RESET button. The following

screen appears.

3. Press and release the AIR ▲ YES button to continue.

If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.

4. The EAGLE 2 will indicate that it is adjusting the zero reading for a few

seconds, then indicate that the operation is complete before returning to
the Calibration Mode Screen.

Performing a Span Adjustment in Auto Calibration

1. Slide the tubing clamp onto the tubing and connect the tubing to the

sample bag’s inlet fitting. Leave the clamp unclamped for now.

PERFORM

AIR ADJUST?

ADJUSTING ZERO

RELEASE AIR BUTTON

ADJUSTING ZERO

COMPLETE

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION