EAGLE 2 Operator’s Manual

ESM-01 Calibration • 191

Calibrating with a 5-Gas Cylinder

The calibration instructions below show a 5 channel instrument which has the
four standard channels, LEL/oxygen/H

2

S/CO, and an ESM-01 channel for

sulfur dioxide (SO

2

).

1. See “Calibration Supplies and Equipment” on page 56 for a description

of the necessary calibration supplies. See Table 19 on page 202 for
available cylinders. Make sure your calibration cylinder is appropriate for
the ESM-01 detection range.

2. Find a fresh-air environment. This is an environment free of toxic or

combustible gases and of normal oxygen content (20.9%).

3. While in Measuring Mode, press and hold the RANGE ▼ SHIFT button,

then press the DISPLAY ADJUST NO button and release both buttons.

4. If the unit prompts you for the password, enter it by using the AIR ▲

YES and RANGE ▼ SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.

5. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

6. Move the cursor to the PERFORM AIR ADJUST menu item by using

the RANGE ▼ SHIFT button.

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION