EAGLE 2 Operator’s Manual

Infrared CO

2

Calibration • 227

5. Put the caps back on the scrubber.

Infrared CO

2

Calibration

An infrared CO

2

channel can be calibrated using the auto calibration method

or the single calibration method. To calibrate an infrared CO

2

channel using

the single calibration method, see “Calibrating Using the Single Calibration
Method” on page 64 and follow the instructions for calibrating a single
channel. If your instrument is a multi-channel instrument that includes one or
more infrared CO

2

channels, RKI Instruments, Inc. recommends using the

auto calibration method for convenience. The calibration instructions below
show a 5 channel instrument which has the four standard channels, LEL/
oxygen/H

2

S/CO, and an infrared CO

2

channel for 0-5 %vol. To use the auto

calibration method to calibrate a multi-channel instrument that includes an
infrared CO

2

channel, do the following:

NOTE: If your instrument has more than one infrared CO

2

sensor, you will

need a calibration cylinder for each sensor.

1. See “Calibration Supplies and Equipment” on page 56 for a description

of the necessary calibration supplies. In addition to an appropriate multi-
gas cylinder that is used to calibrate any active standard channels, you
will also need a cylinder to calibrate the infrared CO

2

channel. See

Table 25 on page 237 for available cylinders. Make sure your calibration
cylinder is appropriate for the infrared CO

2

detection range.

2. Find a fresh-air environment. This is an environment free of toxic or

combustible gases and of normal oxygen content (20.9%).

3. While in Measuring Mode, press and hold the RANGE ▼ SHIFT button,

then press the DISPLAY ADJUST NO button and release both buttons.

4. If the unit prompts you for the password, enter it by using the AIR ▲

YES and RANGE ▼ SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.