EAGLE 2 Operator’s Manual

Using Setup Mode • 125

With LC/BH MODE SELECT set to LEAK CHECK ONLY, the Mode
Select Screen appears when the unit is turned on. You are able to select from
Normal Mode and Leak Check Mode.

With LC/BH MODE SELECT set to OFF, the Mode Select Screen does not
appear when the unit is turned on and the unit goes into Normal Mode after
the start up sequence.

This setting is factory set to OFF when a unit is shipped unless the
instrument is ordered for bar hole measurement or leak checking use. See
“Appendix K: Using the EAGLE 2 in Bar Hole Mode” and “Appendix L:
Using the EAGLE 2 in Leak Check Mode” for discussions of Bar Hole Mode
and Leak Check Mode, respectively.

1. From the main menu, place the cursor next to LC/BH MODE SELECT.

2. Press and release POWER ENTER RESET. The Leak Check/Bar Hole

Mode Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Setting the Bar Hole Measurement Time

This setting indicates the length of time the unit will sample when a bar hole
measurement is initiated in Bar Hole Mode. It can be set to 30 (factory
setting), 45, or 60 seconds.

1. From the main menu, place the cursor next to BH MEASURING TIME.

2. Press and release POWER ENTER RESET. The Bar Hole Measuring

Time Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

LEAK CHECK /

BAR HOLE MODE

UP/DOWN THEN ENTER

OFF

BAR HOLE

MEASURING TIME

UP/DOWN THEN ENTER

30 SECONDS