290 • Start Up and Operation

EAGLE 2 Operator’s Manual

4. The Battery Voltage Screen appears for a few seconds.

5. The Inert Mode Selection Screen will then be displayed.

Use the AIR ▲ YES or RANGE

SHIFT button to display the mode you

wish to enter. To use the EAGLE 2 for tank tester applications, ensure
that the float probe is connected and select Inert Mode. To use the
EAGLE 2 in Normal Mode, select Normal Mode and refer to “Measuring
Mode, Normal Operation” on page 29. If you have not made a selection
after 20 seconds, the instrument will begin to beep and the LEDs will
begin to flash. They will continue until you make a selection.

WARNING: The EAGLE 2 is not a gas monitoring device until you select

a mode and enter it.

6. The warm up sequence will continue as described in “Start Up” on

page 22. The oxygen alarms in the alarm point screens will be the alarm
points for Inert Mode and not for Normal Mode.

7. When the warm up is complete and the instrument is operating in Inert

Mode, the screen will appear as shown below if the instrument is
sampling a fresh air environment. The screen is shown with %LEL and
oxygen channels. Your unit may only have %LEL.

BATTERY MIN:

4.3 VOLTS

BATTERY NOW:

5.2 VOLTS

MODE SELECT

UP/DOWN THEN ENTER

NORMAL MODE

I CH4 0%LEL

N OXY 20.9vol%ALRM2

E

R

T