RKI Instruments EAGLE 2 Manual User Manual

Page 152

Advertising
background image

142 • PID Relative Response Feature

EAGLE 2 Operator’s Manual

Table 15: High Range Relative Response Gas List

The last choice in each list, PID, can be defined by the user and loaded in the
EAGLE 2 using the Eagle 2 Maintenance Data Loader Program. See the
Eagle 2 Maintenance Data Loader Program Operator’s Manual.

3. Use the AIR ▲ YES or RANGE

SHIFT button to move the cursor next

to the desired gas.

4. Press and release POWER ENTER RESET. The PID channel will be

configured to the selected gas and the EAGLE 2 will proceed to the
STEL Screen. This configuration will be in force until either a different
gas is selected in Display Mode or the unit is turned off.

PID

0-50.00

OFF

OFF

OFF

OFF

Target Gas

Detection

Range (ppm)

Alarm 1 Factory

Setting (ppm)

Alarm 2 Factory

Setting (ppm)

STEL

(ppm)

TWA

(ppm)

Acetone

0-1000

500

750

750

500

Benzene

0-1000

50

250

OFF

OFF

Diesel Fuel

NO 1

0-1500

200

300

OFF

200

Ethanol

0-15000

1000

1500

OFF

1000

Gasoline

0-2000

300

500

500

300

Isobutylene

0-2000

400

600

60

42

Isopropanol

0-5000

200

400

400

200

JP-5 Fuel

0-1000

140

210

OFF

14

Methyl Ethyl

Ketone

0-1500

200

300

300

200

Toluene

0-1000

50

150

150

50

N-Hexane

0-5000

500

1000

1000

500

Propylene

0-2500

500

750

OFF

500

Styrene

0-500

20

40

40

20

Tetrachloro-

ethylene

0-1000

25

100

100

25

Trichloro-

ethylene

0-1000

50

100

100

50

Vinyl Chloride

0-4000

100

500

OFF

OFF

PID

0-2000

OFF

OFF

OFF

OFF

Table 14: Low Range Relative Response Gas List