Troubleshooting • 71

• The unit indicates

flow failure and
does not recover
when POWER
ENTER RESET is
pressed and
released.

• The probe tube is

clogged.

• The hydrophobic filter disk

in the probe is dirty.

• The sample hose has a

kink or obstruction.

• The internal hydrophobic

filter is dirty.

• The pump is

malfunctioning.

1. Inspect the probe tube for any

obstructions.

2. Inspect the hydrophobic filter disk

in the probe and replace if
necessary.

3. Inspect the sample hose for kinks

or obstructions and replace if
necessary.

4. Inspect the internal hydrophobic

filter and replace if necessary.

5. If difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• Auto calibration or

single calibration
fails.

• The auto calibration

values may not match the
cylinder gas
concentrations (auto
calibration only).

• The charcoal filter is

saturated causing an
elevated CO reading.

• The sample gas is not

reaching the sensors
because of a bad
connection.

• The calibration cylinder

may be out of gas or is
outdated.

• The sensor for the

affected channel(s) may
need replacement.

1. Check all calibration tubing for

leaks or for any bad connections.

2. Make sure the EAGLE 2 has

been properly set up for
calibration.

3. Change the charcoal filter.
4. Verify that the calibration cylinder

contains an adequate supply of
fresh test sample.

5. If the fail condition continues,

replace the sensor(s).

6. If the difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

• Display indicates

“SYSTEM FAIL
12” during startup.

• A memory error has

occurred.

1. Press and hold the RANGE

SHIFT button, then press the
DISPLAY ADJUST NO button
and release both. The Enter
Password Screen will appear.

2. Enter the password, “1994”, to

proceed to the Set Default
Screen.

3. Press and release the AIR

YES button twice to restore the
defaults. See “Restoring the
Default Settings” on page 128 for
a description of issues to
consider when restoring the
defaults.

4. If difficulties continue, contact

RKI Instruments, Inc. for further
instruction.

Table 9: Troubleshooting the EAGLE 2

Symptoms

Probable Causes

Recommended Action