Parts list, Channel as described in “infrared co, Calibration” on page 227 – RKI Instruments EAGLE 2 Manual User Manual

Page 247

Advertising
background image

EAGLE 2 Operator’s Manual

Parts List • 237

properly, it may be damaged when the top case is re-installed.

16. Make sure that the top case gasket is fully seated in its groove and

carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.

17. Turn the EAGLE 2 upside down and tighten the three case screws to

secure the top case to the bottom case.

18. Calibrate the infrared CO

2

channel as described in “Infrared CO

2

Calibration” on page 227.

Parts List

Table 25: Infrared CO

2

EAGLE 2 Parts List

Part Number

Description

47-5051RK

Infrared CO

2

sensor cable

DEM-3313-1

Infrared CO

2

sensor, 0-5 %vol

DEM-3313-4

Infrared CO

2

sensor, 0-10,000 ppm

DEM-3313-5

Infrared CO

2

sensor, 0-60 %vol

81-0071RK-01

Calibration cylinder, 5000 ppm CO

2

in nitrogen, 34 liter

81-0071RK-03

Calibration cylinder, 5000 ppm CO

2

in nitrogen, 103 liter

81-0072RK-01

Calibration cylinder, 2.5 %vol CO

2

in nitrogen, 34 liter

81-0072RK-03

Calibration cylinder, 2.5 %vol CO

2

in nitrogen, 103 liter

81-0073RK-01

Calibration cylinder, 15 %vol CO

2

in nitrogen, 34 liter

81-0073RK-03

Calibration cylinder, 15 %vol CO

2

in nitrogen, 103 liter

Advertising