RKI Instruments EAGLE 2 Manual User Manual

Page 219

Advertising
background image

EAGLE 2 Operator’s Manual

TC Calibration • 209

7. Press and release the POWER ENTER RESET button. The following

screen appears.

8. Press and release the AIR ▲ YES button to continue.

If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.

9. The EAGLE 2 will indicate that it is adjusting the zero reading for a few

seconds, then indicate that the operation is complete before returning to
the Calibration Mode Screen.

10. Install the demand flow regulator onto the multi-gas calibration cylinder.

11. Connect the sample tubing to the demand flow regulator.

12. Install the probe on the EAGLE 2 inlet fitting. Make sure the probe is

PERFORM

AIR ADJUST?

ADJUSTING ZERO

RELEASE AIR BUTTON

ADJUSTING ZERO

COMPLETE

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION