EAGLE 2 Operator’s Manual

Display Mode • 43

If you do not want to change the setting, press and release the DISPLAY
ADJUST NO button and the unit will return to the Gas Display Screen.

Catalytic (LEL) Sensor Screen

This screen appears only when either a TC sensor or an infrared combustible
sensor is installed in an EAGLE 2 along with a catalytic combustible LEL
sensor. See “Appendix F: TC Sensors”, “Appendix H: Infrared Methane
Sensor”, or “Appendix I: Infrared Hydrocarbon Sensor” for a description of
this screen and instructions to use it.

Methane Elimination Mode Screen

This screen displays only if the EAGLE 2 catalytic combustible channel is
setup for one of the gases in the CONFIGURE GASES menu item in Setup
Mode that supports methane elimination (See “Configuring the Combustible
Gas” on page 108).

The standard setup for methane (CH

4

) does not support methane elimination.

When applicable, use this screen to enable and disable the methane
elimination feature. See “Appendix J: Methane Elimination Mode” for more
discussion of the methane elimination feature.

1. With the Methane Elimination Mode Screen displayed, press and release

the AIR ▲ YES or RANGE

SHIFT button to toggle to the desired

setting, ENABLED or DISABLED.

2. Press and release the POWER ENTER RESET button. The unit will save

the setting and proceed to the next menu item.

If you changed the setting and do not want to save the change, press and
release the DISPLAY ADJUST NO button to continue to the next menu
item without saving the change.

Catalytic Sensor Relative Response Screen

This screen displays only if RELATIVE RESPONSE in Setup Mode is set
to ON (see “Updating the Catalytic Sensor Relative Response Setting” on
page 114). Use this screen to temporarily change the gas configuration of the

METHANE ELIMINATION

MODE

DISABLED