Chapter 4: Calibration Mode

Overview

This section describes the EAGLE 2 in Calibration Mode. In Calibration
Mode, you can move through a menu of screens to do the following:

Perform a span adjustment on all channels simultaneously using
auto calibration

Perform a span adjustment on one channel at a time using single
calibration

Perform a fresh air (zero) adjustment

NOTE: You can set up the EAGLE 2 to alert you during the startup

sequence when calibration is due. See “Updating the Calibration
Reminder Setting” on page 122

CAUTION: BEFORE EACH DAY ’S USAGE, SENSITIVITY IN THE

%LEL RANGE MUST BE TESTED ON A KNOWN
CONCENTRATION OF THE COMBUSTIBLE TARGET GAS,
METHANE, EQUIVALENT TO 25 - 50% OF FULL SCALE
CONCENTRATION (the %LEL full scale is 100 %LEL).
ACCURACY MUST BE WITHIN -0 to + 20% OF ACTUAL.
ACCURACY MAY BE CORRECTED BY FOLLOWING THE
CALIBRATION INSTRUCTIONS FOR THE COMBUSTIBLE
CHANNEL BELOW.

If the combustible channel passes the above response test and does not
require calibration, the unit should still be calibrated periodically. The
optimum frequency of calibration depends heavily on how the EAGLE 2 is
used. For example, instruments used daily may need to be calibrated weekly
or monthly, while instruments that are used only a few times a year may need
to be calibrated before each use. Typical calibration frequencies range from
monthly to quarterly. Make sure to perform the combustible channel response
test as described above and make sure to develop a calibration schedule
tailored to your application that takes this test and required calibration
resulting from this test into account.