234 • Infrared CO

2

Calibration

EAGLE 2 Operator’s Manual

35. If you want to change the infrared CO

2

channel’s calibration gas value,

follow Step 17 - Step 26 above beginning with the infrared CO

2

Calibration Gas Value Screen below instead of the standard channel
Calibration Gas Value Screen.

36. With the infrared CO

2

Calibration Gas Value Screen displayed, press and

release the POWER ENTER RESET button to proceed to the Calibration
In Process Screen for the infrared CO

2

channel with CAL IN PROCESS

flashing.

If you do not want to proceed with the calibration, press and release the
DISPLAY ADJUST NO button to return to the infrared CO

2

Cal Gas

Values Screen.

If you do want to continue with the calibration, proceed to the next step.

37. Screw the infrared CO

2

calibration cylinder onto the demand flow

regulator.

38. Connect the tubing from the regulator to the rigid tube on the probe.

Allow the EAGLE 2 to draw gas for the appropriate time shown in the
table below.

Table 24: IR CO

2

Sensor Calibration Times

39. Press and release the POWER ENTER RESET button to set the span

Range

Calibration Time

0-5.00 %vol

1 minute

0-10,000 ppm

90 seconds

0-60.0 %vol

90 seconds

CAL GAS VALUE

CO2 2.50 vol%

ENTER TO BEGIN CAL

CAL IN PROCESS

CO2 0.00 vol%

ENTER WHEN DONE