Calibrating Using the Auto Calibration Method • 93

17. Fill the gas collection bag by screwing the fixed flow regulator or

dispensing valve onto the calibration cylinder and turning the knob
counterclockwise.

18. Allow the gas to dispense until the gas collection bag is a little over half

full.

19. Turn the knob clockwise to stop the gas flow, clamp down the hose clamp

and remove the regulator or dispensing valve from the cylinder.

20. Disconnect the tubing from the regulator or dispensing valve.

21. Open the hose clamp on the gas bag tubing.

22. Connect the tubing from the gas bag to the rigid tube on the probe. Allow

the gas to flow for one minute.

NOTE: If you are calibrating the catalytic combustible channel with a gas

concentration of 1000 ppm or lower, you must use a 24 inch
humidifier tube to connect the gas bag to the rigid tube on the probe.

23. Press and release the POWER ENTER RESET button to set the span

adjustment to the programmed values.

24. If all channels passed calibration the following screen sequence occurs.

AUTO CALIBRATION

DONE

REMOVE GAS

AUTO CALIBRATION

END

CALIBRATION MODE

> AUTO CALIBRATION
SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION