EAGLE 2 Operator’s Manual Replacing the Hydrophobic Probe’s Particle Filter and Hydro-

6. While the batteries are charging, the green indicator LED will be off and

the amber one will be on.

7. The charging module has an internal timeout feature set at 9.5 hours. A

full charge should be reached in less than 9.5 hours. When a full charge
has been reached, both the green and yellow LEDs will be on.

8. If charging should fail, the green indicator LED will be off and the amber

one will be blinking.

Table 10 summarizes the battery charger conditions.

Replacing the Hydrophobic Probe’s
Particle Filter and Hydrophobic Filter Disk

Inspect the probe’s internal components if you notice that the EAGLE 2’s
pump sounds bogged down or if an unexplained low flow alarm occurs.
Replace the particle filter if it appears to be dirty. Replace the hydrophobic
filter disk if it appears dirty or saturated with liquid. Replace the O-rings in
the probe if either of them appears damaged.

1. Grasp each end of the clear probe body firmly and unscrew the two

halves from each other. One half includes a plastic tube fitting and the
probe tube. The other half includes a metal fitting that mates with the
EAGLE 2 inlet fitting.

Table 10: Battery Charger Conditions

Amber LED

Green LED

Status

ON

OFF

CHARGING

ON

ON

READY/FULL

BLINKING

OFF

FAIL

OFF

ON

CONTINUOUS
OPERATION