36 • Measuring Mode, Alarms

EAGLE 2 Operator’s Manual

Alarm Indications

The EAGLE 2 will sound an alarm and flash the LED arrays when one of the
target gas concentrations rises above the Low Alarm level, or in the case of
oxygen falls below the Low Alarm level, for that gas.

The EAGLE 2 also sounds an alarm and flashes the LED arrays when one of
the target gas concentrations rises above the High Alarm level and when the
STEL and TWA alarm levels are reached for CO and H

2

S.

NOTE: If an alarm condition occurs while you are in Display Mode, the

EAGLE 2 will automatically bring up the alarm screen instead.

When a failure condition occurs, such as a sensor failure, low flow, or dead
battery condition, the unit will also sound an alarm and flash the LED arrays.

The table below summarizes the types of alarms produced by the EAGLE 2
and their indications.

Table 6: Alarm Types and Indications

Alarm Type

Visual Indications

Audible Indication

Low Alarm

Concentration of gas rises
above the Low Alarm setting
or falls below the Low Alarm
setting for O

2

.

ALRM1 appears next to gas

reading

• Alarm LED arrays flash once

per second

• Backlight turns on

Pulsing tone once per
second

High Alarm

Concentration of gas rises
above the High Alarm
setting.

ALRM2 appears next to gas

reading

• Alarm LED arrays flash twice

per second

• Backlight turns on

Pulsing tone twice per
second

TWA or STEL

Concentration of CO or H

2

S

rises above the TWA or
STEL alarm setting.

• Alarm LED arrays flash once

per second

• Backlight turns on

TWA or STEL appears next to

gas reading

Pulsing tone once per
second

Over Range

OVER appears next to gas

reading

• Gas reading indicates full

scale

• Alarm LED arrays flash twice

per second

• Backlight turns on

Pulsing tone twice per
second