224 • Description

EAGLE 2 Operator’s Manual

Figure 38: Typical Infrared CO

2

Sensor Location

The infrared CO

2

sensor is installed in a single sensor flow chamber which is

located in the area next to the standard 4-sensor flow chamber. This area can
accommodate up to three single sensor flow chambers. Figure 38 above
illustrates a typical infrared CO

2

sensor location in front of the pump. The

infrared CO

2

flow chamber may also be installed in one of the other two

sensor chamber locations depending on the particular version of the EAGLE
2. Some infrared CO

2

instrument configurations do not include the 4-sensor

flow chamber.

Infrared CO

2

Sensor

The infrared CO

2

sensor is a cylindrical sensor with a diffusion opening on

the front and a 12 pin circular connector on the back. A 12 wire cable plugs
into the back of the infrared CO

2

sensor with a circular connector that

includes a locking lever. The other end of the cable plugs into an infrared sub
PCB (see description below) that is installed on the main PCB. The sensor is
held in the infrared flow chamber by a bracket on standoffs.

Infrared Sub PCB

The infrared sub PCB is a circuit board that is installed on the main PCB in

H2S Sensor

Pump

Top Case Not Shown

CO2 Scrubber

Infrared CO2 Sensor

Standard Flow Chamber

Charcoal Filter

CO Sensor

Oxygen

Sensor

Bracket

LEL Sensor and
Sensor Bracket

Oxygen Sensor