58 • Calibrating Using the Auto Calibration Method

EAGLE 2 Operator’s Manual

Calibrating Using the Auto
Calibration Method

This method allows you to calibrate the CH

4

(%LEL sensor), oxygen, H

2

S,

and CO sensors simultaneously. It is designed for use with the RKI 4-gas
calibration cylinder and is the quickest and easiest method to calibrate the
EAGLE 2.

Setting the Fresh Air Reading

1. While in the Calibration Mode Screen, move the cursor to the

PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT
button.

2. Press and release the POWER ENTER RESET button. The following

screen appears.

3. Press and release the AIR ▲ YES button to continue.

If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION
> PERFORM AIR ADJUST

NORMAL OPERATION

PERFORM

AIR ADJUST?