256 • Infrared Hydrocarbon Calibration

EAGLE 2 Operator’s Manual

5. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

6. Move the cursor to the PERFORM AIR ADJUST menu item by using

the RANGE ▼ SHIFT button.

7. Press and release the POWER ENTER RESET button. The following

screen appears.

8. Press and release the AIR ▲ YES button to continue.

If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION

PERFORM

AIR ADJUST?