EAGLE 2 Operator’s Manual

Entering Calibration Mode • 57

cylinder. This chapter also includes instructions for calibrating one channel at
a time using single calibration.

Entering Calibration Mode

To enter Calibration Mode, do the following:

1. Find a fresh-air environment. This is an environment free of toxic or

combustible gases and of normal oxygen content (20.9%).

2. While in Measuring Mode, press and hold the RANGE ▼ SHIFT button,

then press the DISPLAY ADJUST NO button and release both buttons.

3. If the unit prompts you for the password, enter it by using the AIR ▲

YES and RANGE ▼ SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.

4. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

NOTE: The following screens illustrate a four-gas EAGLE 2 for detection

of CH

4

(%LEL using catalytic sensor), oxygen, H

2

S, and CO. Your

EAGLE 2 may display slightly different screens.

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION
PERFORM AIR ADJUST

NORMAL OPERATION