EAGLE 2 Operator’s Manual
Maintenance • 169
NOTE: At this point it is recommended that the finger cots be used on the
fingers handling the lamp. Finger cots are included with the lamp
cleaning kit.
10. Hold the PID sensor steady on the working surface with one hand and
using the other hand, locate the tabs on the electrode stack removal tool
and insert them into the slots on the side of the PID sensor near the face.
Figure 33: Using Removal Tool
11. Squeeze the removal tool to push the tabs into the sensor slots until the
electrode stack and lamp are released.
12. Carefully lift the PID sensor body away from the electrode stack and
lamp. If the lamp remains lodged in the sensor body, carefully remove it
with tweezers.
13. If the spring in the lamp cavity comes out, place it back into the lamp
cavity.
14. Discard the old electrode stack.
15. Hold the new electrode stack between the thumb and forefinger of one
hand and place the window end of the lamp inside the O-ring seal in the
new electrode stack with the other hand as shown below. Take care not to
touch the lamp window.