RKI Instruments EAGLE 2 Manual User Manual

Page 169

Advertising
background image

EAGLE 2 Operator’s Manual

PID Calibration • 159

33. If the PID channel passed calibration, the following screen sequence

occurs.

If the PID channel cannot be adjusted to the proper value, a screen displays
that indicates a calibration failure and indicates that the PID sensor failed to
calibrate.

The buzzer and alarm LEDs activate in a double pulsing pattern. Press and
release the POWER ENTER RESET button to reset the alarm and continue to
the Calibration Mode Screen. Attempt to calibrate the PID channel again. If
the failure continues, investigate the cause. See “Troubleshooting” on
page 70

34. Disconnect the tubing from the probe.

35. Unscrew the demand flow regulator from the calibration cylinder.

36. Use the RANGE ▼ SHIFT button to place the cursor next to the

NORMAL OPERATION menu option, then press and release the
POWER ENTER RESET button to return to Measuring Mode.

AUTO CALIBRATION

DONE

REMOVE GAS

AUTO CALIBRATION

END

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

FAILED SENSOR

ENTER TO CONTINUE

<IBL>