EAGLE 2 Operator’s Manual

Start Up, Leak Check Mode • 283

12. The Date/Time Screen appears for a few seconds.

13. If USER/STATION ID is turned on (see “Turning the User/Station ID

Function On or Off” on page 118), the ID Screen appears for a few
seconds.

If USER/STATION ID is turned off, only the serial number is shown.

14. If the EAGLE 2 experiences a catalytic combustible sensor failure during

start up, a screen indicating which sensor failed appears and the buzzer
sounds a pulsing tone twice per second. In the example below, the H

2

S

sensor has failed.

If the catalytic combustible sensor fails, it is not possible to enter
Leak Check Mode. Press and release the POWER ENTER RESET
button to acknowledge the failure and return to the Mode Select
Screen. Replace the failed sensor.

If any other sensor that is installed fails, press and release the
POWER ENTER RESET button to acknowledge the failure and
continue to Leak Check Mode. Change the failed sensor(s) as soon
as possible for use in Normal Mode.

9/12/2008

15:00:00

USER ID

MIKE

STATION ID

PUMP 1

SERIAL NUMBER

E2A515

FAILED SENSOR(S)

<CH4> < >

< > < >

ENTER TO CONTINUE