EAGLE 2 Operator’s Manual

Start Up • 25

If the unit is due for calibration and CAL PAST DUE ACT is set to
CONFIRM TO CAL, then the following screen displays and the
buzzer sounds in a double pulsing pattern.

To perform a calibration, press and release the AIR ▲ YES button.
The EAGLE 2 will enter Calibration Mode and the LCD will show
the Calibration Mode main menu. See “Chapter 4: Calibration Mode”
on page 55 for instructions to calibrate the EAGLE 2. When you are
done with the calibration and exit Calibration Mode, the unit will
begin the startup sequence. If the calibration was successful, the
screen above will not appear again until the unit is due for
calibration. If the calibration was not successful, the screen above
will again appear in the startup sequence.

To continue without performing a calibration, press and release the
DISPLAY ADJUST NO button.

If the unit is due for calibration and CAL PAST DUE ACT is set to
MUST CALIBRATE, then the following screen displays and the
buzzer sounds in a double pulsing pattern.

The EAGLE 2 cannot be used until a successful calibration has been
performed. Press and release the ENTER button to enter Calibration
Mode. See “Chapter 4: Calibration Mode” on page 55 for
instructions to calibrate the EAGLE 2.

NOTE: In this situation, even if the password function has been turned on,

no password is required to perform a calibration.

When you are done with the calibration and exit Calibration Mode,
the unit will begin the startup sequence. If the calibration was
successful, the screen above will not appear again until the unit is due
for calibration. If the calibration was not successful, the screen above

CALIBRATION DATE

IS PAST DUE

PERFORM

CALIBRATION?

CALIBRATION DATE

IS PAST DUE

ENTER TO PERFORM

CALIBRATION