66 • Calibrating Using the Single Calibration Method

EAGLE 2 Operator’s Manual

4. Move the cursor next to the SINGLE CALIBRATION menu item by

using the AIR ▲ YES button.

5. Press and release the POWER ENTER RESET button. The Select Sensor

Screen appears with the cursor flashing.

6. Move the cursor next to the sensor you want to calibrate with the AIR ▲

YES and RANGE ▼ SHIFT buttons. In the example below, the CH

4

sensor is selected for span adjustment.

If you do not want to proceed with the span adjustment, press and release
the DISPLAY ADJUST NO button or place the cursor next to ESCAPE
and press and release POWER ENTER RESET to return to the
Calibration Mode Screen.

To proceed with the calibration, continue with the next step.

CALIBRATION MODE

AUTO CALIBRATION

> SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

SELECT SENSOR
TO CALIBRATE

>ESCAPE

CH4 OXY
H2S CO

SELECT SENSOR
TO CALIBRATE

ESCAPE

>CH4 OXY

H2S CO