EAGLE 2 Operator’s Manual

Description • 205

Figure 37: Typical TC Sensor Location

TC Sensor

The TC sensor’s appearance is exactly the same as that of the LEL sensor. Its
housing includes a sintered metal flame arrestor on one end that allows gas to
diffuse into the sensor. On the other end, five pins extend from the sensor.
The TC sensor can be distinguished from the LEL sensor by the part number
imprinted on it. The TC sensor part number is TE-7568 while the LEL part
number is NC-6260B. The sensor cable connects to pins on one end and
terminates in a four-position connector on the other end which plugs into the
TC sub PCB (see description below). The sensor bracket is installed over the
TC sensor to keep it seated in place.

NOTE: The LEL and TC sensors and sensor cables are identical in

appearance. Take care not to plug the LEL sensor cable into the port
on the TC sub PCB and not to plug the TC sensor cable into the LEL
port on the main PCB.

TC Sub PCB

The TC sub PCB is a circuit board that is installed on the main PCB in one of
the 3 sub PCB positions when a TC sensor is used with the EAGLE 2. The

CO Sensor

Pump

LEL Sensor and
Sensor Bracket

Charcoal Filter

Oxygen Sensor
Bracket

Standard Flow Chamber

Top Case Not Shown

TC Sensor

H2S Sensor

Oxygen Sensor