124 • Using Setup Mode

EAGLE 2 Operator’s Manual

2. Press and release POWER ENTER RESET. The Calibration Past Due

Action Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating the Calibration Interval

This setting defines the amount of time between calibrations. The time can be
set in 1 day increments. The minimum setting is 1 day and the maximum
setting is 365 days. The factory setting is 90 days.

1. From the main menu, place the cursor next to CAL INTERVAL.

2. Press and release POWER ENTER RESET. The Set Calibration Interval

Screen appears.

3. Use AIR ▲ YES or RANGE ▼ SHIFT to display the desired setting.

Tip: Press and hold AIR ▲ YES or RANGE ▼ SHIFT to rapidly scroll

through the settings.

4. Press and release POWER ENTER RESET to save the setting and return

to the main menu.

Updating LC/BH Mode Setting

With LC/BH MODE SELECT set to LC & BH, the Mode Select Screen
appears when the unit is turned on. You are able to select from Normal Mode,
Leak Check Mode, and Bar Hole Mode.

With LC/BH MODE SELECT set to BAR HOLE ONLY, the Mode Select
Screen appears when the unit is turned on. You are able to select from
Normal Mode and Bar Hole Mode.

CALIBRATION PAST

DUE ACTION

UP/DOWN THEN ENTER

CONFIRM TO CAL

SET CALIBRATION

INTERVAL

UP/DOWN THEN ENTER

90 DAYS