EAGLE 2 Operator’s Manual

Configuring the PID Gas in Setup Mode • 175

12. Use AIR ▲ YES and RANGE ▼ SHIFT to place the cursor next to the

desired gas. In the example below, the cursor is placed next to
isopropanol (IPA).

Table 16: PID Gas Configurations

Gas Name (LCD Abbreviation)

Detection Range Full Scale

Low Range/High Range

Acetone (ACT)

30.00 ppm/1000 ppm

Benzene (BNZ

25.00 ppm/1000 ppm

Diesel Fuel No. 1 (DSL)

40.00 ppm/1500 ppm

Ethanol (ETA)

400.0 ppm/15000 ppm

Gasoline (GSL)

50.00 ppm/2000 ppm

Isobutylene (IBL)
Standard Factory Gas Configuration

50.00 ppm/2000 ppm

Isopropanol (IPA)

200.0 ppm/5000 ppm

JP-5 Fuel (JP5)

30.00 ppm/1000 ppm

Methyl Ethyl Ketone (MEK)

40.00 ppm/1500 ppm

Toluene (TOL)

25.00 ppm/1000 ppm

N-Hexane (HEX)

200.0 ppm/5000 ppm

Propylene (PRL)

50.00 ppm/2500 ppm

Styrene (STY)

20.00 ppm/500 ppm

Tetrachloroethylene (PCE)

30.00 ppm/1000 ppm

Trichloroethylene (TCE)

30.00 ppm/1000 ppm

Vinyl Chloride (VCM)

100.0 ppm/4000 ppm

PID (PID)
User Defined Gas Loaded with
Maintenance Program

50.0 ppm/2000 ppm

PID(0- 2000ppm IBL )

2000 ppm GSL

2000 ppm IBL

> 5000 ppm IPA

1000 ppm JP5