RKI Instruments EAGLE 2 Manual User Manual

Page 223

Advertising
background image

EAGLE 2 Operator’s Manual

TC Calibration • 213

27. Press and release the POWER ENTER RESET button to set the span

adjustment for each channel to the programmed values.

28. If all channels passed calibration, the following screen sequence occurs.

If any of the sensors cannot be adjusted to the proper value, a screen displays
that indicates a calibration failure and lists the sensor(s) that failed to
calibrate. In the example below, the oxygen and H

2

S channels failed

calibration. The other sensors calibrated normally.

The buzzer and alarm LEDs activate in a double pulsing pattern. Press and
release the POWER ENTER RESET button to reset the alarm and continue to
the Calibration Value Screen for the TC channel. After calibrating the TC
channel by following the instructions below, attempt to calibrate the standard
channels again. If the failure continues, investigate the cause. See
“Troubleshooting” on page 70

29. Remove the tubing from the rigid tube on the probe.

30. Unscrew the 4-gas cylinder from the demand flow regulator.

AUTO CALIBRATION

DONE

REMOVE GAS

AUTO CALIBRATION

END

CAL GAS VALUE

CH4 50.0 %vol.

ENTER TO BEGIN CAL

FAIL SENSOR

< > <OXY>
<H2S> < >

ENTER TO CONTINUE

Advertising