218 • Configuring the TC Gas in Setup Mode

EAGLE 2 Operator’s Manual

4. The “S” will then disappear and the following screen will appear for a

few seconds.

5. If the unit prompts you for the password, enter it by using AIR ▲ YES

and RANGE ▼ SHIFT to select each password number and then pressing
and releasing the POWER ENTER RESET button to enter it and move on
to the next number until all of the numbers are entered. The main menu
displays. It displays six menu items at a time.

6. Use the RANGE ▼ SHIFT button to move the cursor down the menu to

CONFIGURE GASES.

7. Press and release POWER ENTER RESET. The Configure Gases Screen

appears with the cursor flashing next to CAT, the catalytic sensor. If an
EAGLE 2 has a TC sensor installed, one of the three optional sensor
types, OP1, OP2, or OP3 will indicate it is a TC. In the example below,
OP1 is shown as a TC sensor.

GAS MONITOR

EAGLE 2

>SET DATE & TIME

SET DATE FORMAT

SET BATTERY TYPE

CONFIGURE CHANNELS
CONFIGURE GASES

CATALYTIC UNITS

CONFIGURE GASES

> CAT : CH4 (CAT)

OP1 : CH4 (TC)

OP2 : --- (---)

OP3 : --- (---)