EAGLE 2 Operator’s Manual

Start Up, Leak Check Mode • 281

10. The gas alarm setpoints are displayed by three screens in sequence: the

Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen.
Each screen remains on the LCD for three seconds.

11. After the alarm screens, if CAL REMINDER is turned on, the screen that

appears next depends on how CAL PAST DUE ACT is set in the Setup
Mode Menu (see “Updating the Calibration Past Due Action Setting” on
page 123).

If the unit is due for calibration and CAL PAST DUE ACT is set to
CONFIRM TO CAL, then the following screen displays and the
buzzer sounds in a double pulsing pattern.

To perform a calibration, press and release the AIR ▲ YES button.
The EAGLE 2 will enter Calibration Mode and the LCD will show
the Calibration Mode main menu. See “Chapter 4: Calibration Mode”
on page 55 for instructions to calibrate the EAGLE 2. When you are

A CH4 50 %LEL

HL OXY 23.5 vol%

I A H2S 30.0 ppm

GR CO 50 ppm

HM

S

CH4 50 vol%

A CH4 10 %LEL

L L OXY 19.5 vol%

O A H2S 10.0 ppm

WR CO 25 ppm

M

S

CH4 10 vol%

ALARMS STEL & TWA

H2S(ppm) 15.0 10.0

CO (ppm) 200 25

CH4(vol%) OFF OFF

CALIBRATION DATE

IS PAST DUE

PERFORM

CALIBRATION?