202 • Parts List

EAGLE 2 Operator’s Manual

the ESM-01 chamber.

21. Install the two sensor bracket screws.

22. Confirm that the main PCB is seated in its slots and that its bottom edge

is resting on the bottom of the bottom case. If the main PCB is not seated
properly, it may be damaged when the top case is re-installed.

23. Make sure that the top case gasket is fully seated in its groove and

carefully put the top case back on the bottom case. If you have any
difficulty mating the top and bottom cases, inspect the placement of the
main PCB and the placement of the top case gasket.

24. Turn the EAGLE 2 upside down and tighten the three case screws to

secure the top case to the bottom case.

25. Calibrate the ESM-01 channel as described in “ESM-01 Calibration” on

page 181.

Parts List

Table 19: ESM-01 EAGLE 2 Parts List

Part Number

Description

33-7120RK

H

2

S scrubber disk, for ESM-01DH-D-HCN sensor

33-7121RK

H

2

S scrubber disk, for ESM-01DH-F-SO2 sensor

47-5015RK

ESM-01 sensor cable

ESM-01DH-ASH3

Arsine sensor, 0-1.50 ppm

ESM-01R-NH3

Ammonia sensor, 0-75.0 ppm

ESM-01DH-D-HCN

Hydrogen cyanide sensor, 0-15.0 ppm

ESM-01DH-PH3

Phosphine sensor, 0-1.00 ppm

ESM-01DH-F-SO2

Sulfur dioxide sensor, 0-6.00 ppm

ESM-K01-CL2

Chlorine sensor, 0-3.00 ppm

81-0142RK-02

Calibration cylinder, 5-gas (SO

2

, CH

4

, O

2

, H

2

S, CO), 58-liter aluminum

81-0142RK-04

Calibration cylinder, 5-gas (SO

2

, CH

4

, O

2

, H

2

S, CO), 34-liter aluminum

81-0170RK-02

Calibration cylinder, 5 ppm SO

2

in nitrogen, 58 liter

81-0170RK-04

Calibration cylinder, 5 ppm SO

2

in nitrogen, 34 liter

81-0175RK-02

Calibration cylinder, 10 ppm NH

3

in nitrogen, 58 liter

81-0175RK-04

Calibration cylinder, 10 ppm NH

3

in nitrogen, 34 liter