RKI Instruments EAGLE 2 Manual User Manual

Page 304

Advertising
background image

294 • Alarms

EAGLE 2 Operator’s Manual

3.

The LCD will show the following screen for a few seconds with
the “S” in the lower right corner indicating the unit is entering
Setup Mode.

4.

The “S” will then disappear and the following screen will appear
for a few seconds.

5.

If the unit prompts you for the password, enter it by using the AIR
▲ YES or RANGE ▼ SHIFT buttons to select each password
number and then pressing and releasing the POWER ENTER
RESET button to enter it and move on to the next number until all
of the numbers are entered. The main menu displays. It displays
six menu items at a time.

6. Use the AIR ▲ YES or RANGE ▼ SHIFT button to move the cursor to

the ALARM POINTS menu item.

7. Press and release POWER ENTER RESET. The Change Alarm Point

Settings Screen appears and all detection channels are displayed.

GAS MONITOR

EAGLE 2 S

GAS MONITOR

EAGLE 2

>SET DATE & TIME

SET DATE FORMAT

SET BATTERY TYPE

CONFIGURE CHANNELS
CONFIGURE GASES

CATALYTIC UNITS

CHANGE ALARM

POINT SETTINGS

> 1: CH4 2: OXY

3: H2S 4: CO

END