228 • Infrared CO

2

Calibration

EAGLE 2 Operator’s Manual

5. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

6. Move the cursor to the PERFORM AIR ADJUST menu item by using

the RANGE ▼ SHIFT button.

7. Remove the black caps from the ends of the carbon dioxide scrubber. Be

sure to grab the scrubber by the EAGLE 2 fitting so that the fitting and
tubing do not come off.

8. Attach the CO

2

scrubber to the inlet fitting of the EAGLE 2 and allow the

instrument to draw gas for 1 minute before performing an air adjust.

NOTE: Attaching the CO

2

scrubber to the inlet fitting eliminates the

background CO

2

found in fresh air and allows the EAGLE 2 to

obtain an accurate zero reading.

9. Press and release the POWER ENTER RESET button. The following

screen appears.

10. Press and release the AIR ▲ YES button to continue.

If you do not want to continue, press the DISPLAY ADJUST NO button
and the unit will return to the Calibration Mode Screen.

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION

> PERFORM AIR ADJUST

NORMAL OPERATION

PERFORM

AIR ADJUST?