Using setup mode – RKI Instruments EAGLE 2 Manual User Manual
Page 113

EAGLE 2 Operator’s Manual
Using Setup Mode • 103
•
Press the DISPLAY ADJUST NO button while in a screen where you are
entering or updating parameters to exit the screen without saving any
changes.
Using Setup Mode
WARNING: The EAGLE 2 is not in operation as a gas detector while in
Setup Mode.
1. Take the EAGLE 2 to a non-hazardous location and turn it off if it is on.
2. Press and hold the AIR ▲ YES and RANGE ▼ SHIFT buttons, then
press and hold the POWER ENTER RESET button. When you hear a
beep, release the buttons.
3. The LCD will show the following screen for a few seconds with the “S”
in the lower right corner indicating the unit is entering Setup Mode.
4. The “S” will then disappear and the following screen will appear for a
few seconds.
GAS MONITOR
EAGLE 2 S
GAS MONITOR
EAGLE 2