258 • Infrared Hydrocarbon Calibration

EAGLE 2 Operator’s Manual

13. Move the cursor next to the AUTO CALIBRATION menu item by

using the AIR ▲ YES button.

14. Press and release the POWER ENTER RESET button to display the

Calibration Gas Values Screen.

The gas concentrations displayed in the Calibration Gas Values Screen
must match the gas concentrations listed on the 4-gas calibration cylinder.
If all concentrations match, go to Step 25.
If one or more concentrations do not match, continue with Step 15.
If you do not want to continue with the calibration, press and release the
DISPLAY ADJUST NO button to return to the Calibration Mode Screen.

NOTE: The RKI 4-gas cylinder typically contains 12% O

2

by volume.

When using the auto calibration method, be sure to set the “OXY”
auto calibration value to agree with the concentration listed on the
cylinder’s label, not zero.

15. To adjust the values on the screen, press and hold the RANGE ▼ SHIFT

button, then press the DISPLAY ADJUST NO button and release both.
The following screen appears with the cursor next to CH4.

16. Place the cursor next to the channel whose gas value you want to change

using the AIR ▲ YES and RANGE ▼ SHIFT buttons.

17. Press and release the POWER ENTER RESET button to select the

CALIBRATION MODE

> AUTO CALIBRATION

SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

CAL GAS VALUES
CH4 50 %LEL
OXY 12.0 vol%
H2S 25.0 ppm

CO 50 ppm

ENTER TO BEGIN CAL

ADJUST AUTO

CALIBRATION VALUES

> CH4 50 %LEL

OXY 12.0 vol%
H2S 25.0 ppm

CO 50 ppm