88 • Calibrating Using the Auto Calibration Method

EAGLE 2 Operator’s Manual

3. If the unit prompts you for the password, enter it by using the AIR ▲

YES and RANGE ▼ SHIFT buttons to select each password number and
then pressing and releasing POWER ENTER RESET to enter the number
and move on to the next one.

4. The Calibration Mode Screen displays with the cursor next to AUTO

CALIBRATION.

NOTE: The following screens illustrate a 4-gas EAGLE 2 for detection of

CH

4

(%LEL using catalytic sensor), oxygen, H

2

S, and CO. Your

EAGLE 2 may display slightly different screens.

Calibrating Using the Auto
Calibration Method

This method allows you to calibrate the CH

4

(%LEL catalytic combustible

sensor), oxygen, H

2

S, and CO sensors simultaneously. It is designed for use

with the RKI 4-gas calibration cylinder and is the quickest and most
convenient method to calibrate the EAGLE 2.

Setting the Fresh Air Reading

1. While in the Calibration Mode Screen, move the cursor to the

PERFORM AIR ADJUST menu item by using the RANGE ▼ SHIFT
button.

CALIBRATION MODE

> AUTO CALIBRATION
SINGLE CALIBRATION

PERFORM AIR ADJUST

NORMAL OPERATION

CALIBRATION MODE

AUTO CALIBRATION

SINGLE CALIBRATION
> PERFORM AIR ADJUST

NORMAL OPERATION