Select user id screen – RKI Instruments EAGLE 2 Manual User Manual

Page 58

Advertising
background image

48 • Display Mode

EAGLE 2 Operator’s Manual

4. Press and release the AIR ▲ YES button to scroll through screens that

display the rest of the active channels’ alarm settings.

Select User ID Screen

This screen displays only if USER/STATION ID in the Setup Mode menu is
set to ON (see “Turning the User/Station ID Function On or Off” on
page 118). Use this screen to select a user ID from the user ID list in the
EAGLE 2’s memory. The current user ID is displayed. A user ID can be up to
16 characters long. The EAGLE 2 can store up to 32 user IDs.

The user ID provides a way to identify the EAGLE 2 user during a data
logging session. If the user ID is changed during an operating session, a new
data session is initiated with the new user ID attached to it. This allows you to
change the user ID during operation and have each user ID that was used
during an operating session saved for the corresponding data. See the Eagle 2
Data Logger Management Program Operator’s Manual for a detailed
description of data logging and the user ID.

The user ID list cannot be edited using the EAGLE 2 user interface. The
Eagle 2 Maintenance Data Loader Program is required to define or change
user IDs in the user ID list. For a detailed description of editing the list of

CHANNEL 2

OXY 0- 40.0 vol%
LO ALRM: 19.5 vol%

HI ALRM: 23.5 vol%

CHANNEL 3

H2S 0-100.0 ppm

LO ALRM: 5.0 ppm
HI ALRM: 30.0 ppm

STEL ALRM: 5.0 ppm

TWA ALRM: 1.0 ppm

CHANNEL 1

CH4 0- 100 %LEL

LO ALRM: 10 %LEL
HI ALRM: 50 %LEL

LO ALRM: 5000 ppm
HI ALRM:25000 ppm

CHANNEL 4

CO 0- 500 ppm

LO ALRM: 25 ppm
HI ALRM: 50 ppm

STEL ALRM: 200 ppm

TWA ALRM: 25 ppm

AIR

YES

AIR

YES

AIR

YES

AIR

YES

SELECT

USER ID?

CURRENT USER ID
JOHN