166 • Maintenance

EAGLE 2 Operator’s Manual

using the other hand, locate the tabs on the electrode stack removal tool
and insert them into the slots on the side of the PID sensor near the face.

Figure 31: Using Removal Tool

11. Squeeze the removal tool to push the tabs into the sensor slots until the

electrode stack and lamp are released.

12. Carefully lift the PID sensor body away from the electrode stack and

lamp. If the lamp remains lodged in the sensor body, carefully remove it
with tweezers.

13. If the spring in the lamp cavity comes out, place it back into the lamp

cavity.

14. Discard the old PID lamp.

NOTE: At this point it is recommended that the finger cots be used on the

fingers handling the lamp. Finger cots are included with the lamp
cleaning kit.

15. Hold the electrode stack between the thumb and forefinger of one hand

and place the window end of the new lamp inside the O-ring seal in the
electrode stack with the other hand as shown below. Take care not to
touch the lamp window.